Translate this page into:
An unusual case of guide wire fracture during percutaneous transluminal coronary angioplasty
*For correspondence: sezenbaglan@hotmail.com
-
Received: ,
This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
This article was originally published by Medknow Publications & Media Pvt Ltd and was migrated to Scientific Scholar after the change of Publisher.
A 64 yr old female patient† was referred to the department of Cardiology, Faculty of Medicine, Ufuk University, Ankara, Turkey, with chest pain, in September 2016. A drug-eluting stent (DES) was implanted in the left anterior descending artery (LAD) in another centre a month ago. Coronary angiogram of the left system showed a fractured remnant guide wire extending from the first diagonal ostium to the thoracic aorta (Figs. 1, 2 and Video). It was decided to catch the distal tip of the remnant wire with loop snare catheter and to retrieve the remnant wire by deep-guide catheter wedge technique (Video). The remnant wire fractured by several attempts and turned to a spiral shape in the proximal part of LAD (Fig. 3). Broken wire parts were removed from the circulatory system. Since this technique failed, by passing through the spiral shape fractured remnant wire, a 3.0 mm×18 mm DES was implanted in the LAD urgently. The mass of the remnant wire was crushed onto the vessel wall (Video). The retained guide wire tip was isolated from the circulation, and post-procedural TIMI (thrombolysis in myocardial infarction) 3 flow was achieved in the LAD (Video). One month later, control coronary angiography was performed and observed normal intracoranary blood flow (Video).

- An angiographic image of the fractured wire shows an entrapped guide wire in the left anterior descending. The arrow shows proximal tip of the remnant guide wire.

- The angiography shows the fractured wire in the left anterior descending artery, and the arrows show the extension of the same to the ascending aorta.

- Spiral shape of the remnant fractured guide wire in the left anterior descending artery.